1 Departments of Mechanical Engineering and Cardiothoracic Surgery, Washington University St. Louis, Missouri 63130
In a continuing effort to develop improved prosthetic heart valves, a redefinition of the anatomy of the human aortic valve as a function of stress was undertaken. Dimensions and geometric relationships of the human aortic valve as a function of intraaortic pressure between 0 and 120 mm Hg were obtained from a series of silicone rubber valve casts. The axial length of the valve region was found to vary negligibly with pressure, but significant variations in geometry and angular dimensions were seen. The leaflet attachment annulus forms an ellipse at the plane of intersection with the cylindrical surface passing from the left ventricular tract through the aorta. Deductions from stress considerations for the measured geometry indicate that the loaded leaflet is a section of a cylindrical surface. The equation for this developed surface was obtained, and a prosthetic design was determined using average values at 100 mm Hg. The leaflet is developable onto a plane with a cut required along part of the junction line between the initially cylindrical part and the plane coapting surfaces. Optimum valve shape mandates a base angle between the cylindrical leaflet and the center axis of 70° (
=20-22°, where
is the leaflet angle).
Submitted on January 31, 1974
Accepted on August 8, 1974
This article has been cited by other articles:
![]() |
A. C. Kollar, S. D. Lick, D. M. Palacio, and R. F. Johnson Jr Ross procedure with a composite autograft using stretch gore-tex material. Ann. Thorac. Surg., October 1, 2009; 88(4): e34 - e36. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Katayama, N. Umetani, S. Sugiura, and T. Hisada The sinus of Valsalva relieves abnormal stress on aortic valve leaflets by facilitating smooth closure. J. Thorac. Cardiovasc. Surg., December 1, 2008; 136(6): 1528 - 1535. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Piazza, P. de Jaegere, C. Schultz, A. E. Becker, P. W. Serruys, and R. H. Anderson Anatomy of the Aortic Valvar Complex and Its Implications for Transcatheter Implantation of the Aortic Valve Circ Cardiovasc Interv, August 1, 2008; 1(1): 74 - 81. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David, C. M. Feindel, G. D. Webb, J. M. Colman, S. Armstrong, and M. Maganti Aortic Valve Preservation in Patients With Aortic Root Aneurysm: Results of the Reimplantation Technique Ann. Thorac. Surg., February 1, 2007; 83(2): S732 - S735. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-J. Schafers, D. Aicher, F. Langer, and H. F. Lausberg Preservation of the Bicuspid Aortic Valve Ann. Thorac. Surg., February 1, 2007; 83(2): S740 - S745. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David, C. M. Feindel, G. D. Webb, J. M. Colman, S. Armstrong, and M. Maganti Long-term results of aortic valve-sparing operations for aortic root aneurysm. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 347 - 354. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-J. Schafers, B. Bierbach, and D. Aicher A new approach to the assessment of aortic cusp geometry. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 436 - 438. [Full Text] [PDF] |
||||
![]() |
A. Ranga, O. Bouchot, R. Mongrain, P. Ugolini, and R. Cartier Computational simulations of the aortic valve validated by imaging data: evaluation of valve-sparing techniques Interactive CardioVascular and Thoracic Surgery, August 1, 2006; 5(4): 373 - 378. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. A. Goetz, K. H. Lim, R. Ibled, N. Grousson, S. L. H. Salgues, and J. H. Yeo Forces at single point attached commissures (SPAC) in pericardial aortic valve prosthesis Eur. J. Cardiothorac. Surg., February 1, 2006; 29(2): 150 - 155. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Thubrikar, M. R. Labrosse, K. J. Zehr, F. Robicsek, G. G. Gong, and B. L. Fowler Aortic root dilatation may alter the dimensions of the valve leaflets Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 850 - 855. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. De Paulis, C. Schmitz, R. Scaffa, P. Nardi, L. Chiariello, and H. Reul In vitro evaluation of aortic valve prosthesis in a novel valved conduit with pseudosinuses of Valsalva J. Thorac. Cardiovasc. Surg., October 1, 2005; 130(4): 1016 - 1021. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. G. Gleason New graft formulation and modification of the David reimplantation technique J. Thorac. Cardiovasc. Surg., August 1, 2005; 130(2): 601 - 603. [Full Text] [PDF] |
||||
![]() |
R. Cartier and M. Chaput Annulus Sizing in Aortic Remodeling Procedure Ann. Thorac. Surg., December 1, 2004; 78(6): 2215 - 2216. [Full Text] [PDF] |
||||
![]() |
F. Robicsek, M. J. Thubrikar, and A. A. Fokin Cause of degenerative disease of the trileaflet aortic valve: review of subject and presentation of a new theory Ann. Thorac. Surg., April 1, 2002; 73(4): 1346 - 1354. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. C. Gasparyan METHOD OF DETERMINATION OF AORTIC VALVE PARAMETERS FOR ITS RECONSTRUCTION WITH AUTOPERICARDIUM: AN EXPERIMENTAL STUDY J. Thorac. Cardiovasc. Surg., February 1, 2000; 119(2): 386 - 387. [Full Text] [PDF] |
||||
![]() |
P. Dagum, G. R. Green, F. J. Nistal, G. T. Daughters, T. A. Timek, L. E. Foppiano, A. F. Bolger, N. B. Ingels Jr, and D. C. Miller Deformational Dynamics of the Aortic Root : Modes and Physiologic Determinants Circulation, November 9, 1999; 100 (2009): II-54 - II-62. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Choo and C. M.G. Duran A surgical method for selecting appropriate size of graft in aortic root remodeling Ann. Thorac. Surg., February 1, 1999; 67(2): 599 - 599. [Full Text] [PDF] |
||||
![]() |
K. Morishita, T. Koshino, and T. Abe A surgical method for selecting appropriate size of graft in aortic root remodeling Ann. Thorac. Surg., February 1, 1999; 67(2): 599 - 600. [Full Text] [PDF] |
||||
![]() |
C. M. G. Duran, B. Gometza, N. Kumar, R. Gallo, and R. Martin-Duran AORTIC VALVE REPLACEMENT WITH FREEHAND AUTOLOGOUS PERICARDIUM J. Thorac. Cardiovasc. Surg., August 1, 1995; 110(2): 511 - 516. [Abstract] [Full Text] |
||||
![]() |
T. R. Zwink, I. G. Burwash, C. Y. Miyake-Hull, and C. M. Otto Changes in Aortic Annulus Diameter During the Cardiac Cycle and its Effect on Predicting Aortic Valve Prosthesis Size Journal of Diagnostic Medical Sonography, September 1, 1994; 10(5): 262 - 267. [Abstract] [PDF] |
||||
![]() |
G.W. Christie Anatomy of aortic heart valve leaflets: the influence of glutaraldehyde fixation on function Eur. J. Cardiothorac. Surg., January 1, 1992; 6(suppl_1): S25 - S33. [Abstract] [PDF] |
||||
|
Circulation Research Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1974 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |