Articles |
Strong Children's Research Center, Department of Pediatrics, University of Rochester (NY) School of Medicine and Dentistry.
Correspondence to Masaaki Yoshigi, MD, University of Rochester, Department of Pediatrics, 601 Elmwood Ave, Box 631, Rochester, NY 14642. E-mail myoshigi@cvdev.rochester.edu.
| Abstract |
|---|
|
|
|---|
Key Words: embryonic impedance linearity frequency domain cycle length
| Introduction |
|---|
|
|
|---|
Hydraulic energy is imparted during systole from the heart and dissipated as blood courses through the body. A portion of this energy is pulsatile and distends blood vessels while the remaining energy produces forward blood flow. Pulsatile energy may be a potential promoter of vascular development. Pulsatile pressure-flow relationships are also critical determinants of vascular load that may influence cardiac morphogenesis. The quantitative analysis of pulsatile pressure-flow relationships provides insight into the structure and regulation of the cardiovascular system.6
The calculation of vascular impedance by Fourier methods and Windkessel-type lumped parameter models is based on the assumption that pressure and flow are linearly related in the frequency domain. Linearity in the frequency domain can be referred to as "independence of harmonics," which means that one harmonic is not influenced by the other harmonics. Independence of harmonics ensures a unique relationship between corresponding harmonics of pressure and flow and, therefore, justifies the calculation of vascular impedance and hydraulic energy. This assumption of linearity has been tested in the mature pulmonary and systemic circulation and forms the basis of our understanding of arterial afterload.7 8 9 10 11 12
The direct proof of linearity was obtained in vitro by driving an isolated vascular system with a pure sinusoidal flow input and then determining pulsatile pressure-flow relationships with the use of Fourier analysis.7 Linearity in vivo was first reported by Bergel and Milnor,9 who studied the effects of changing the cycle length for the pulmonary circulation. In response to changes in cycle length, impedance spectra were relatively constant, and the scatter of the harmonics was acceptably small. Thus, they concluded that the pulmonary circulation was "approximately" linear on the basis of the qualitative consideration. This finding, in a highly compliant pulmonary circulation, actually conflicts with the nonlinear properties of arterial pressure-flow relations. In a strict sense, if arterial compliance was related to cycle length, there should have been some changes in curvature in the impedance spectrum and some scatter, particularly in the intermediate frequency range. Nevertheless, this method to test linearity in vivo was soon applied to the systemic circulation,10 11 12 and a similar conclusion was drawn. Thus, the use of cycle length change has been accepted as a valid, though imperfect, method to test the linearity of pressure-flow relations in the frequency domain.6
In the present study, we characterize the hemodynamic factors previously reported to be involved in the regulation of morphogenesis, growth, and function of the embryonic cardiovascular system.13 Although arterial impedance has been calculated in the embryonic circulation,1 14 the assumption of linearity in the embryonic vascular system has not been proven. Therefore, we tested the linearity of the chick embryonic vasculature by changing cycle length without altering peripheral vascular resistance. We found that the stage 24 embryonic chick vascular system can be approximated as a linear system from 0 to 6 Hz, and this finding supports our analytic paradigm using Fourier analysis and lumped parameter modeling.
| Materials and Methods |
|---|
|
|
|---|
0 (angular frequency). The two parameters were calculated as described elsewhere16 and determined with various sizes of the pipette tip. The size of the pipette tip was quantified by the resistance of the pipette at a current of 50 µA. The correlations between the resistance (R, in M
) and these two parameters were as follows: ß=0.11·R+0.18 (r=.69, P<.0001) and
0=-4.4·R+12.7 (r=.72, P<.001), with R ranging from 0.2 to 0.6 M
. We measured the resistance of each pipette before the individual experiments and corrected Fourier terms of the pressure in the analysis process.16 Dorsal aortic blood flow velocity was measured simultaneously with a 20-MHz pulsed-Doppler velocimeter (model 545-C, University of Iowa). A 0.5-mm-diameter Doppler crystal was positioned at a 45° angle to the aorta using a protractor jig at the same level as the pressure pipette. The angle was corrected in the analysis process. The Doppler velocimeter focus range is 0 to 10 mm, with a sample volume of 1 to 4 mm.17 This sample volume includes the entire dorsal aortic cross section (0.41±0.01 mm in diameter, stage 24).1 Pressure and velocity analog waveforms were continuously digitized at a sampling rate of 500 Hz and analyzed using a custom virtual instrument (LabVIEW, National Instruments).
We also measured the internal diameter of the dorsal aorta using videomicroscopy. A calibration scale scribed with 50-µm divisions was positioned perpendicular to the long axis of the dorsal aorta. At the current level of optical resolution, we noted no change in aortic diameter during the normal and altered cardiac cycles. Therefore, we determined a single value for mean dorsal aortic diameter. Assuming aortic cross section to be circular, instantaneous aortic blood flow (Q) was calculated using the following equation: Q=
·d2·V/4, where d is the mean aortic diameter and V is the instantaneous velocity.
Experimental Protocols
Baseline data were recorded for at least 15 seconds, and intrinsic cycle length was determined. We then changed cycle length using a 1-mm-diameter steel probe preheated (hot probe alone group, n=12) or precooled (cold probe alone group, n=11) in a water bath.18 19 The probe was applied to the sinus venosus for 3 seconds. Cycle length decreased in response to the hot probe and increased in response to the cold probe. Only embryos that recovered to 90% to 110% of intrinsic cycle length 30 seconds after the intervention were included for analysis. In a separate set of embryos, the alternate probe was applied after recovery to intrinsic cycle length from the first intervention (hot-and-cold group, n=15). In the hot-and-cold group, a hot probe was applied first in nine embryos and was followed by a cold probe; the order was reversed in six embryos. Sham (control) embryos were treated by the probe, heated to 37°C to 38°C. All experiments were performed within 180 seconds.
Data Analysis and Statistics
The cycle length of most embryos recovered to 90% to 110% of its intrinsic cycle length within 50 to 60 beats (20 to 30 seconds) after probe application. Therefore, we chose three consecutive waveforms at baseline and then immediately, 10 seconds, and 20 seconds after the probe application. Cycle length, mean pressure, pulse pressure, mean blood flow, stroke volume, and mean vascular resistance at each time point were determined by averaging the three consecutive cycles. Similarly, the impedance spectrum at each time point was determined by averaging three corresponding harmonics with the use of statistics for circular data.20
The calculation of input impedance spectrum is published elsewhere.1 6 Experimental impedance spectra with varying cycle lengths were superpositioned on the same axes to identify the scatter of the harmonics along a reference curve. Since each spectrum has a zero-order modulus at 0 Hz, we averaged all moduli at 0 Hz to determine the single value at 0 Hz. All harmonics above 12 Hz were discarded because of the limited frequency response of the servonull pressure system. Superpositioned impedance spectra were thus determined for individual embryos. The number of harmonics up to 12 Hz varied among groups, because the impedance spectrum derived from longer cycle lengths has more harmonics than from shorter cycle lengths.
Hemodynamic parameters in the time domain were summarized as mean±SEM and compared with matched sham (control) values by repeated measures ANOVA. The significance of the order of interventions in the hot-and-cold group was also tested by repeated measures ANOVA. The response to individual probe in the hot-and-cold group was compared with corresponding probe alone by two-way repeated measures ANOVA. Statistical significance was defined by a value of P<.05. Calculations were performed with SPSS (SPSS Inc) and Statistica (Statsoft) software.
Linearity of Pulsatile Pressure-Flow Relations
We evaluated the scatter of the harmonics along a reference curve as a quantitative measure of the linearity in the frequency domain. The reference curve was derived from the three-element Windkessel (3-WK) model, which consists of a proximal resistor in series with a parallel arrangement of a distal resistor and a capacitor.21 Biological correlates of the proximal and distal resistors and the capacitor are the characteristic impedance (Rc), peripheral resistance (Rp), and total arterial compliance (C), respectively. Model impedance Zm(j
) of the 3-WK model is given in rectangular form by the following:
![]() | (E1) |
is complex angular frequency. For individual embryos, we optimized three lumped parameters using modified Levenberg-Marquardt nonlinear regression. To calculate the regression, the experimental impedance spectrum, Ze(j
), was expressed by rectangular form:
![]() | (E2) |
) and Zim(
) are the real and imaginary components of the harmonics at frequency
, respectively. The residual sum of squares, SSres, was defined as follows22 :
![]() | (E3) |
![]() | (E4) |
and
are means of the real and imaginary components, respectively. Then, the coefficient of determination, R2, as a measure of "goodness of fit" was calculated by (SStot-SSres)/SStot. We applied Fisher's z transformation and F test to compare the difference in R2 among groups with different interventions. In addition, we compared the difference in the scatter of harmonics among three frequency ranges (0 to 4, 4 to 8, and 8 to 12 Hz). SSres/df (where df indicates degrees of freedom) was calculated for individual frequency ranges in the hot-and-cold group embryos (n=15) and assessed by F test. The harmonic at 0 Hz was eliminated from this analysis.
Distribution of Hydraulic Energy Over the Frequency Range
In the mature vascular system, the ratio of pulsatile energy to total hydraulic energy is considered to be a reverse index of energy efficiency.23 Of note, the pulsatile ratio is higher in the embryonic vascular system than in the mature circulation,14 indicating that pulsatile energy likely plays a significant role in the maturation process. To assess physiological relevance of linearity, distribution of the hydraulic energy over the frequency range was also calculated as pulsatile energy, Wpuls(
), distributed from >0 Hz to
Hz (note that 0 Hz is not included)23 :
![]() | (E5) |
)|, |Z(j
)|, and
(
) are the modulus of flow, the modulus of impedance, and the phase of impedance, respectively. The highest value of
was set to 12 Hz so that "total pulsatile" energy was defined by Wpuls(12). Steady state energy, W(0) (because it is energy at 0 Hz), was calculated from the product of mean pressure and mean flow. Then, hydraulic energy, W(
), distributed from
0 Hz to
Hz (note that 0 Hz is included) was defined by W(0)+Wpuls(
). Hence, total hydraulic energy was defined by W(12). Then, percentage ratios of Wpuls(
) to total pulsatile energy Wpuls(12) and W(
) to total hydraulic energy W(12) were calculated for individual embryos. The calculations were performed for the baseline data in the hot-and-cold group (n=15). | Results |
|---|
|
|
|---|
|
|
Impedance Spectra With Varying Cycle Lengths
When we varied cycle length by a single intervention, experimental impedance spectra were consistent, and most of the harmonics up to 6 or 7 Hz fell on the same curve (Fig 2
). Model impedance spectra with optimized lumped parameters are also superpositioned in the figures. Experimental impedance spectra in response to sequential intervention (hot-and-cold group) and fitted model impedance are shown in Fig 3
. A larger number of harmonics were obtained with the sequential intervention. In all groups, we noted minimal scatter of the impedance data along the reference curve from 0 to 6 Hz, although there was larger scatter in the higher frequency range. The experimental impedance phase spectrum had a zero crossover between 6 and 8 Hz. In addition, difference between experimental and model impedance in the phase spectrum was larger in the higher frequency range. Table 2
summarizes optimized lumped parameters and the parameters for "goodness of fit." The overall coefficient of determination (R2) was near unity. The parameters for "goodness of fit" were similar among groups regardless of the type of interventions (P>.07 for Fisher's z, by F test). Fig 4
shows comparison among three different frequency ranges in terms of the scatter along the model impedance. F test revealed larger scatter in the higher frequency range (8 to 12 Hz) compared with lower frequency ranges.
|
|
|
|
Distribution of Hydraulic Energy in the Frequency Range
Percentage ratios of Wpuls(
) to total pulsatile energy, Wpuls(12), increased from 4.2±1.66% at 2 Hz to 68.5±1.13% at 4 Hz, 85.0±0.76% at 6 Hz, 93.7±0.35% at 8 Hz, and finally 97.4±0.23% at 10 Hz. Percentage ratios of W(
) to total hydraulic energy, W(12), increased from 73.1±0.56% at 0 Hz to 74.3±0.59% at 2 Hz, 91.5±0.3% at 4 Hz, 96.0±0.18% at 6 Hz, 98.3±0.1% at 8 Hz, and finally 99.3±0.07% at 10 Hz.
| Discussion |
|---|
|
|
|---|
The linearity test in vivo depends on the evidence that vascular properties, particularly peripheral vascular properties, do not change in response to alterations in cycle length.9 In the present study, this evidence was confirmed, since mean vascular resistance was unchanged in all groups. As shown in Figs 2 and 3![]()
, our data indicate that the vascular system is linear from 0 to 6 or 7 Hz by the small variance around the reference curve. However, because of some scatter in the impedance spectrum with increasing frequency, we propose that the in vivo embryonic vascular system should be considered quasilinear or conditional linear.
As described by previous investigators,8 24 25 "nonlinearity" of pulsatile pressure-flow relations has been shown in vascular systems. Several factors, including flow turbulence in the vessels, shear forces due to the viscous properties of blood, and nonlinear pressure-volume relations of the arterial wall, have been inferred as sources of nonlinearity. Among these, Stergiopulos et al24 used a completely controlled computer model and found that the scatter in the impedance spectrum was mainly due to the elastic nonlinearity of the arterial wall. In fact, as Stergiopulos et al pointed out, even the impedance spectrum shown to be linear in vivo9 10 displays scatter, particularly in the intermediate to higher frequency ranges. In our results, particularly some embryos in the cold probe group, scatter already starts around 5 Hz, as seen in Fig 2
. Although statistical difference between the hot probe group and the cold probe group in terms of scatter was not demonstrated, this observation may reflect changes in pressure-volume relations of the arterial wall that are possibly due to decreases in mean pressure, particularly seen in the cold probe group. Recognizing that any instantaneous changes in cycle length, pressure, and flow can alter pressure-volume relations of the arterial wall, linearity is valid only when the focus of the analysis is a first approximation. In addition to these physiological factors, measurement errors are the most likely to cause the scatter at higher frequencies, especially with the low frequency response of our manometer system.
Universality and Limitation of the 3-WK Model
It is of interest that the same 3-WK model, developed for the elastic adult aorta, could accurately describe the impedance characteristics of the embryonic vascular system with an aortic diameter <0.5 mm and 1/70 of the adult mean pressure. Since the pressure waveforms exhibit diastolic pressure, Windkessel capacitance and discharge characteristics of the artery, which deliver steady flow to the periphery from intermittent cardiac output, are already present in the early embryonic circulation. These functional similarities between mature and embryonic circulations may exist because the embryonic vascular system already possesses qualitative characteristics similar to those of the mature vascular system, whereas the difference in scale discriminates between the two systems. Provided that the vascular system is composed of larger proximal vessels, branching vessels, and smaller distal vessels, numerous studies have shown that the pulsatile arterial pressure-flow relationships can generally be characterized by the three parameters Rc, Rp, and C (from Equation 1) at least within the low to mid frequency ranges.26 This may apply to the embryonic circulation, since the structure of the embryonic vascular system itself is similar in terms of the basic components. In addition, although speculative, material properties of the embryonic arterial wall are likely matched relative to the pressure range to achieve the Windkessel function of the artery. Thus, the developmental changes in in vitro material properties of the embryonic arterial wall deserve further investigations.
Although the 3-WK model provided a good fit to the experimental impedance spectrum in the present study, further refinements can be made. As shown in Fig 3
, the superpositioned impedance phase tended to have a zero crossover at 6 or 7 Hz, and the modulus tended to have a minimum at 5 or 6 Hz. These tendencies, summarized as fluctuation of the impedance spectrum, indicate significant wave reflection phenomena in the actual arterial system6 that cannot be reproduced by the 3-WK model. The 3-WK model as a reference curve was suitable for the present study because of its simplicity. However, simple electrical analog models assume an infinite pulse-wave velocity and therefore do not account for wave propagation/reflection phenomena. Ideally, to account for reflected waves, distributed models or asymmetrical T-tube models are required.27 Still, refinement of the electrical analog model might be helpful because of its simplicity of calculations and biological relevance of those lumped parameters. One possibility is proximal addition of an inductance term to the 3-WK model. This type of four-element model can reproduce phase-zero crossover and at least one minimum of the modulus. Actually, parallel addition of an inductance term improved R2 from .966±.006 to .976±.007 (Fig 5
) in the preliminary study. However, quantitative comparisons among various types of analog models and consideration of overparameterization are required to appropriately characterize the embryonic vascular system. Moreover, biological relevance of the inductance term should be examined if the inductance is necessary and sufficient.
|
Differences Between Embryonic and Mature Circulations
The ratio Rp/Rc has been reported to be consistent in the mature systemic circulation and range from 16.7 to 23.9 in most mammals.26 Therefore, the impedance spectrum normalized by Rc represents a fundamental shape of the spectrum for most mammals. Since Rc is one of the two determinants (Rc and C) of the pulsatile load in the arterial system, Rp/Rc is inversely correlated to pulsatile ratio of the hydraulic energy in the arterial system. As calculated from the data in Table 2
, Rp/Rc is
5.5 to 6.8 in the embryonic circulation. The smaller Rp/Rc in the embryonic circulation is consistent with a higher pulsatile ratio of the hydraulic energy and may particularly indicate stiffness of the proximal artery. In addition, greater pulse pressure relative to pressure range indicates that the entire embryonic arterial system is less compliant than the mature circulation. Cross-species comparison of arterial compliance is possible when the time constant (
) of the diastolic pressure decay, a product of Rp and C, is normalized by a duration of the diastolic time (Td).28 Smaller animals with shorter
have shorter Td values to prevent diastolic pressure from decaying to insufficient coronary perfusion pressure, whereas larger animals with longer
have longer Td values. Therefore,
/Td is constant among mature animals and ranges from 3.1 to 6.2, and the optimal heart rate for the species is thus set depending on the
value. The embryonic
/Td is 1.7 to 1.8, calculated from data previously published1 and in Table 2
. Again, the smaller
/Td in the early embryonic circulation supports a less compliant arterial system in the embryo. It is of note that this less compliant system with a relatively low diastolic pressure precedes coronary arterial connections, which are established on embryonic day 7 or 8.29 Developmental changes in pulsatility of the pressure waves, arterial compliance, and pulsatile hydraulic energy should be correlated to the development in coronary circulation.
Study Implications
Despite the technical limitations mentioned above, the linearity of embryonic pressure-flow relations proven here from 0 to 6 Hz in the frequency domain is important new information, given the dramatic differences in scale, histology, and regulation between developing and mature circulations. Even restricted to 6 Hz, a linear system analysis provides valuable hemodynamic information, because most hydraulic energy, 85.0% of the pulsatile and 96.0% of the total energy, is distributed within this frequency range. Based on this linear relationship, it is justified to superimpose impedance spectra with various cycle lengths in order to improve frequency resolution (maximize the number of harmonics). This improvement in the frequency resolution allows a more accurate assessment of arterial compliance and characteristic impedance, especially given the limitation of the manometer's frequency response. Quantitative assessment shown here for the linear/nonlinear issue may provide further significant information because it enables quantitative comparisons among different species, developmental stages, perturbed hemodynamic conditions, and various types of analog models.
Thus, despite the dramatic differences in scale between the embryonic and mature arterial vasculatures, the stage 24 embryonic chick vascular system can be approximated as a linear system from 0 to 6 Hz. Using these techniques, we should be able to investigate wave propagation/reflection phenomena, which are likely to be physiologically important in the rapidly expanding embryonic vasculature.
| Acknowledgments |
|---|
| Footnotes |
|---|
Received February 1, 1996; accepted June 11, 1996.
| References |
|---|
|
|
|---|
2. Burke RD, Wang D, Jones VM. Ontogeny of vessel wall components in the outflow tract of the chick. Anat Embryol. 1994;189:447-456.[Medline] [Order article via Infotrieve]
3. Rosenquist TH, McCoy JR, Waldo KL, Kirby ML. Origin and propagation of elastogenesis in the developing cardiovascular system. Anat Rec. 1988;221:860-871.[Medline] [Order article via Infotrieve]
4. Duband JL, Gimona M, Scatena M, Sartore S, Small JV. Calponin and SM 22 as differentiation markers of smooth muscle: spatiotemporal distribution during avian embryonic development. Differentiation. 1993;55:1-11.[Medline] [Order article via Infotrieve]
5. Hughes AFW. The histogenesis of the arteries of the chick embryo. J Anat. 1942;77:266-287.
6. Nichols WW, O'Rourke MF. McDonald's Blood Flow in Arteries: Theoretical, Experimental, and Clinical Principles. 3rd ed. Philadelphia, Pa: Lea & Febiger; 1990.
7. Attinger EO, Anne A, McDonald DA. Use of Fourier series for the analysis of biological systems. Biophys J. 1966;6:291-304.
8.
Dick DE, Kendrick JE, Matson GL, Rideout VC. Measurement of nonlinearity in the arterial system of the dog by a new method. Circ Res. 1968;22:101-111.
9.
Bergel DH, Milnor WR. Pulmonary vascular impedance in the dog. Circ Res. 1965;16:401-415.
10.
O'Rourke MF, Taylor MG. Vascular impedance of the femoral bed. Circ Res. 1966;18:126-139.
11.
Noble MI, Gabe IT, Trenchard D, Guz A. Blood pressure and flow in the ascending aorta of conscious dogs. Cardiovasc Res. 1967;1:9-20.
12.
Nichols WW, Conti CR, Walker WE, Milnor WR. Input impedance of the systemic circulation in man. Circ Res. 1977;40:451-458.
13. Keller BB. Overview: functional maturation and coupling of the embryonic cardiovascular system. In: Clark EB, Markwald RR, Takao A, eds. Developmental Mechanisms of Heart Disease. New York, NY: Futura; 1995:367-385.
14.
Zahka KG, Hu N, Brin KP, Yin FCP, Clark EB. Aortic impedance and hydraulic power in the chick embryo from stages 18 to 29. Circ Res. 1989;64:1091-1095.
15. Hamburger V, Hamilton HL. A series of normal stages in the development of the chick embryo. J Morphol. 1951;88:49-92.
16. Milnor WR. Methods of measurement. In: Hemodynamics. 2nd ed. Baltimore, Md: Williams & Wilkins; 1989:294-334.
17.
Quail AW, Cottee DBF, White SW. Limitations of a pulsed Doppler velocimeter for blood flow measurement in small vessels. J Appl Physiol. 1993;75:2745-2754.
18.
Cuneo B, Hughes S, Benson DW Jr. Heart rate perturbation in chick embryos: a comparison of two methods. Am J Physiol. 1991;260:H1864-H1869.
19.
Casillas CB, Tinney JP, Keller BB. Influence of acute alterations in cycle length on ventricular function in chick embryos. Am J Physiol. 1994;267:H905-H911.
20. Fisher NI. Statistical Analysis of Circular Data. New York, NY: Cambridge University Press; 1993.
21. Westerhof N, Bosman F, De Vries CJ, Noordergraaf A. Analog studies of the human systemic arterial tree. J Biomech. 1969;2:121-143.
22.
Grant BJ, Paradowski LJ. Characterization of pulmonary arterial input impedance with lumped parameter models. Am J Physiol. 1987;252:H585-H593.
23.
Milnor WR, Bergel DH, Bargainer JD. Hydraulic power associated with pulmonary blood flow and its relation to heart rate. Circ Res. 1966;19:467-480.
24.
Stergiopulos N, Meister JJ, Westerhof N. Scatter in input impedance spectrum may result from the elastic nonlinearity of the arterial wall. Am J Physiol. 1995;269:H1490-H1495.
25. Fung YC. Theoretical pulmonary microvascular impedance. Ann Biomed Eng. 1972;1:221-245.[Medline] [Order article via Infotrieve]
26.
Westerhof N, Elzinga G. Normalized input impedance and arterial decay time over heart period are independent of animal size. Am J Physiol. 1991;261:R126-R133.
27.
Campbell KB, Burattini R, Bell DL, Kirkpatrick RD, Knowlen GG. Time-domain formulation of asymmetric T-tube model of arterial system. Am J Physiol. 1990;258:H1761-H1774.
28. Westerhof N, Elzinga G. Why smaller animals have higher heart rates. Adv Exp Med Biol. 1993;346:319-323.[Medline] [Order article via Infotrieve]
29.
Poelmann RE, Gittenberger-de Groot AC, Mentink MMT, Bokenkamp R, Hogers B. Development of the cardiac coronary vascular endothelium, studied with antiendothelial antibodies, in chicken-quail chimeras. Circ Res. 1993;73:559-568.
This article has been cited by other articles:
![]() |
J. L. Lucitti, R. Visconti, J. Novak, and B. B. Keller Increased arterial load alters aortic structural and functional properties during embryogenesis Am J Physiol Heart Circ Physiol, October 1, 2006; 291(4): H1919 - H1926. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Tobita, E. A. Schroder, J. P. Tinney, J. B. Garrison, and B. B. Keller Regional passive ventricular stress-strain relations during development of altered loads in chick embryo Am J Physiol Heart Circ Physiol, June 1, 2002; 282(6): H2386 - H2396. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. B. Keller, M. Yoshigi, and J. P. Tinney Ventricular-vascular uncoupling by acute conotruncal occlusion in the stage 21 chick embryo Am J Physiol Heart Circ Physiol, December 1, 1997; 273(6): H2861 - H2866. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Chabert and L. A. Taber Intramyocardial pressure measurements in the stage 18 embryonic chick heart Am J Physiol Heart Circ Physiol, April 1, 2002; 282(4): H1248 - H1254. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Tobita, E. A. Schroder, J. P. Tinney, J. B. Garrison, and B. B. Keller Regional passive ventricular stress-strain relations during development of altered loads in chick embryo Am J Physiol Heart Circ Physiol, June 1, 2002; 282(6): H2386 - H2396. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Research Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1996 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |