Articles |
From the Center for Bioengineering, University of Washington, Seattle.
Correspondence to James B. Bassingthwaighte, MD, PhD, Center for Bioengineering, WD-12, University of Washington, Seattle, WA 98195. E-mail jbb@nsr.bioeng.washington.edu.
| Abstract |
|---|
|
|
|---|
and the outflow as
t-
-1, where
is interpreted to be
the dimensionless exponent of a fractal power law relation
characterizing the self-similarity inherent in each curve. (2) The
fractional escape rate, given by the outflow curve divided by the
residue curve, diminished almost exactly as
t-1. In 18 sets of curves,
averaged
2.21±0.27. These results lead to an improved method for extrapolating
the downslopes of indicator dilution curves to estimate their areas and
therefore the blood flows. The evidence also points strongly to the
conclusions that myocardial water washout is a fractal process and that
stirred tank models are inappropriate for the heart.
Key Words: flow-limited blood-tissue exchange power law kinetics positron emission oxygen-15 capillary permeability statistical self-similar processes
| Introduction |
|---|
|
|
|---|
, equals V/F, where
V (in milliliters per gram) is the volume of distribution of
the tracer within the organ, and F is the flow (in
milliliters per gram per minute). This is simply a statement of
conservation of mass and is true whether or not there are limitations
to the exchange rate by membrane barriers or slow diffusion.
Organs are not normally homogeneously perfused. Regional
flows in the heart are broadly heterogeneous.3
In both the heart4 and the lung,5 the spatial
variation appears to be fractal. This means that the spatial
heterogeneity shows statistical self-similarity.
The apparent variance or standard deviation of the regional flows is
dependent on spatial resolution: observing smaller regions with higher
resolution reveals further variation. Finding that the smaller regions
are nonuniform is proof that the overall apparent
heterogeneity must increase with higher resolution. In
special cases, a proportional increase in resolution reveals a
proportional increase in heterogeneity. This relation
follows the power law function:
![]() | (1) |
Turn now to the temporal characteristics or kinetics of tracer exchange
in the heart. These are not predictable from a knowledge of regional
spatial flow distributions. For example, the spatial flow distributions
are approximately gaussian: assuming that the local volume of
distribution, Vi, for a tracer is the
same in all regions, then the time course of tracer transit times for
an intravascular tracer traveling between inflow and outflow would be
directly calculable. The mean transit time for the ith path,
i, is as follows:
![]() | (2) |
is would be skewed to the right in a
fashion dependent on the distribution of
Fis.
However, there is some heterogeneity of regional blood
volumes,6 7 8 so the analysis of washout times would
be further confounded by the volume variation in an unpredictable
fashion. The alternative is to seek a flow-limited marker with a
relatively uniform distribution space, ie, one with a steady state
volume of distribution that has very small regional variation. The
fractional water content of the myocardium is remarkably
uniform7 9 10 at 0.78±0.01 mL/g. The washout curves for
water, and for the lipid-soluble substance antipyrine, have been
shown to be flow-limited in the heart.11 Therefore,
these are the preferred test tracers for examining the shapes of
washout curves. These washout curves will not be influenced by
variation in Vi or by any diffusional
retardation, nor will they be skewed by slow barrier penetration, such
as that which prolongs the tails of dilution curves for potassium or
sucrose or larger hydrophilic solutes. Consequently, washout curves for
water will be governed by the velocities through the hundreds of
thousands of vessels of the arterial and venous networks
and scaled through the local equilibration via capillary-tissue
exchange by the ratios of blood space to total water space. The washout
time course will thus be dominated by pathway transit time
distributions; the influence of the blood-tissue exchange process
for water is to render all the Vis uniform,
leaving the
is reciprocal to the local
Fis. The observations of Rose et
al12 suggest a small barrier limitation in the highest
flow regions, which might affect the upslope and peak of the curves but
would not affect the tails of the curves or our analyses.
This analysis focuses on the question of the nature of the distribution of washout times. The resultant observation that the distribution appears fractal is revealed for the first time in the present study. The result contrasts remarkably with and clearly refutes the long-held perception that washout is an exponential process. This idea was the basis of the Stewart-Hamilton method13 for extrapolating the tails of dilution curves monoexponentially in order to measure the areas under indicator dilution curves for the estimation of organ flow or cardiac output. The results to be shown below will be used to argue that the Stewart-Hamilton method slightly but systematically underestimates the areas and so gives a minor degree of overestimation of flows.
| Materials and Methods |
|---|
|
|
|---|
121 seconds.) The perfusate was Krebs-Ringer bicarbonate
solution with 1% albumin, to which was added triply washed
beef red blood cells to a hematocrit of 20%. The partial pressure of
oxygen was 145 to 180 mm Hg. The isolated heart was mounted on a
perfusion apparatus. The residual tracer activity within
the heart, R(t), was obtained via a pair of NaI
crystal gamma detectors situated on opposite sides of the heart and
used in coincidence mode to provide a measure of the two
simultaneous gamma emissions resulting from positron
annihilation events occurring in the heart. The coincidence method
reduces the counting of emissions from other sources, including scatter
from outside the cylindrical region in which the heart was situated.
The coronary sinus and right ventricular thebesian
veins drained into an outflow cannula inserted through the
pulmonary valve. The weight of blood in the outflow cannula,
positioned below heart level, holds the pressure in the right ventricle
at a negative pressure; therefore, the ventricle stays empty.
15O detectors sensitive directly to the positron (positive
beta) particles were set up on the inflow tubing and outflow tubing,
giving the inflow, Cin(t), and
outflow concentration-time curves, C(t). The
structure of the detectors (made in our laboratory) was similar in
design to that described by Lerch et al.14 All three
curves were corrected for the isotopic decay of 15O, which
has a half-life of 121 seconds. By this experimental approach, we attempted to minimize error in the data acquisition. Because the outflow curve is a slightly dispersed and delayed measure of the derivative of the residue curve and because both outflow and residue curves must conserve mass (account for the amount of tracer in the inflow curve), analyzing the set of curves together minimizes error due to incompleteness of outflow collection because of leaks or contamination of R(t) by any accumulations of tracer leaked from the heart and included in the signal because of the poor collimation.
Analytical Methods and Indicator Dilution Theory
Basic conservation theory, as summarized by
Zierler,15 maintains that after an impulse injection at
t=0 into the entrance to an organ, the fractional residual
content R(t) at the next instant is 100% of the
injectate. R(t) remains at 1.0 until there is an
appearance of tracer in the outflow. The fraction of the injectate
appearing in the outflow per unit time is h(t)
(fraction per second). The accumulated outflow is the integral of the
outflow response and is the total dose, unity, minus the fraction of
dose still retained within the organ, R(t):
![]() | (3) |
is a dummy variable used in the integration.
Further, in the ideal situation, in the absence of error in the data
acquisition, the outflow fraction of the dose appearing per second is
the derivative of the organ content:
![]() | (4) |
![]() | (5) |
is the mean flow, and the asterisk
denotes the convolution integration, more formally written as
Q(t)=
0tCin(
) · R(t-
)d
.
Likewise, the outflow concentration-time curve,
C(t), is as follows:
![]() | (6) |
![]() | (7) |
![]() | (8) |
, so that
h(t)=
C(t)/qo,
illustrating that h(t) is the fraction of the
dose reaching the outflow per unit time.
The fractional escape rate,
(t), is defined as the
fraction of residual tracer escaping per unit time, and is therefore
given directly by the amount of tracer appearing in the outflow per
unit time,
C(t), where
is the total flow, divided by the amount of
tracer retained within the organ at each time t:
![]() | (9) |
|
A Fractal Theory To Be Tested
The fundamental feature of a fractal is self-similarity or
self-affinity. A fractal washout process is therefore one for which
the rate of washout decreases by some exact proportion for some chosen
proportional increase in time; the self-affinity requirement is
fulfilled whenever the "exact proportion" remains unchanged,
independent of the moment or the segment of the data set selected to
measure the proportionality constant. The length of time or the portion
of the washout curve for which this relation holds is known as the
"scaling region" and is necessarily finite, as is the case for
all natural, nonmathematical, fractals. Fractal washout behavior can
begin after the input to the organ is complete. If complete, then one
would expect the late portion of R(t) to have the
shape of a power law relation defining the proportionality
relation:
![]() | (10) |
is the power law exponent. (The observation of a power
law relation is not an explanation of the phenomenon but provides an
incentive for a search for an explanation.) This power law relation
defines the log-log slope
dlogR/dlogt=-
. The
corollary of this is that the outflow tracer concentration-time
curve should also be fractal and since
h(t)=-dR/dt, then taking
the derivative of the right side of Equation 10
![]() | (11) |
-1.
According to the hypothesis, expressed in Equations 10 and 11,
(t) should be a power law function with an exponent equal
to -1:
![]() | (12) |
be used to fit Equations 10 and 11 to the
data simultaneously obtained for R(t)
and h(t)? If so, then does Equation 12
(t),
with the specific power law slope of -1; ie, does
(t)
decay at a rate proportional to 1/time? | Results |
|---|
|
|
|---|
Fig 1
shows the type of curves obtained. Both the residue and the
outflow curves are continuously concave upward on the semilogarithmic
plot (left); therefore, these curves are not
monoexponential. On the log-log plot, both
R(t) and h(t) are
apparently straight for times >
1 minute. The slope of the outflow
tracer concentration-time curve is steeper than that of the residue
function, as was seen for all of the sets of data; this is a major
point, for it denies the suitability of a stirred tank or mixing
chamber model where R(t) and
h(t) would have the same slope fitted by a single
exponential.
Because the input to the system is somewhat dispersed rather than being an ideal instantaneous impulse, the values of R(t) do not peak at t=0; the peaks of R(t) occurred at 8 to 30 seconds, averaging 19±8 seconds, whereas the peaks of h(t) occurred at 14 to 47 seconds, averaging 29±12 seconds. The delays in the tubing from the heart to the outflow detector are on the order of 1 to 3 seconds. (The volume was 0.12 mL, and the flows were 5 to 30 mL/min.) The dispersion due to this tubing has no detectable influence on the shape or slope of the outflow curves.
The data of highest relevance to our analysis are the
tails of the curves beyond the first minute, where the washout is
clearly not significantly influenced by the form of the input, because
all of the transit times are long compared with the duration of the
input. The early parts of R(t) and of
h(t) are therefore clearly influenced by the form
of the input function, but the later portions of the curves are not,
since the input was complete in 10 to 15 seconds. The subsequent
analysis is performed on the data beyond, where
R(t)/Rmax<0.2; these are
the data beyond the first 50 to 80 seconds after the injection, and the
data in most runs extend to
300 seconds.
Unconstrained Fitting of R, h, and

The tails of the residue R(t), outflow
h(t), and the escape rate
(t) for
two experiments are plotted on log-log scales in Fig 2
. Each curve was fit with the log-log regression to
give best estimates of the power law exponent. The values from 18
experiments for
from the residue functions, fitting Equation 10
,
were 2.12±0.33 (mean±1 SD); the values for
from the outflow
curves, fitting Equation 11
, were 2.13±0.34. These values were
obtained independently from residue and outflow data. The coefficients
of variation for the fits to the R(t)s averaged
0.073±0.042 (N=18); the coefficients of variation for the fits to the
outflow curves averaged 0.074±0.070 (N=18). Thus, the independently
estimated values of
from residue and outflow were close to each
other (2.12 and 2.13). The paired differences averaged 0.005, with a
standard deviation of 0.43 for the 18 pairs. Thus, the first two tests
listed below Equation 12
are satisfied and do not cause us to reject
the fractal relation, since the two independent estimates of
are
not statistically different, as tested by either unpaired or paired
differences by Student's t test.
|
The curves for the fractional escape rate
(t) are
necessarily noisier than those for either h(t) or
R(t), since
(t) is calculated by
dividing one curve by the other (Equation 9
). Because of the noise, the
coefficients of variation for the power law regression fitting are
larger (0.21±0.16, N=18). The power law exponents obtained for
unconstrained fitting of the equation
(t)=kt-ß gave estimates of
ß=1.12±0.47 (N=18), so that the average of the unconstrained
estimates of ß did not differ statistically from 1.0, the value
theoretically anticipated in Equation 12
. Again, the fractal hypothesis
cannot be rejected.
Comparing Power Law With Monoexponential
Fits
Fig 3
shows optimized best fits using both the
power law expressions,
(t)=a1t-
1
and
h(t)=a2
2t-
2-1,
and the exponential expressions,
(t)=A1e-k1t
and
h(t)=A2e-k2t.
The test was designed as a test of appropriateness of extrapolation to
predict the shape of the last parts of the tails of
R(t) and h(t); to do this,
the model functions were fit to only a limited segment of the data: for
R(t), from 0.2 to 0.1 only; for
h(t), over exactly the same time period as for
R(t). For this test, the power law fits were not
constrained to use the same
for R(t) and
h(t) but were best-fitting regressions,
log-log for the power law and log-linear for the exponential.
The rate constants for the exponential fits represent best fits
over the same segments of R(t) and
h(t) as were used for the power law fitting.
|
The two sets of data shown in Fig 3
on semilogarithmic, not
log-log, plots are representative of the predictive
capacity of the two extrapolation methods. The power law extrapolation
beyond the time where R(t)=0.1 always lies above
the exponential best fit for both R(t) and
h(t). For 15 of 18 R(t)s,
the power law extrapolation predicted and fit the data for
R(t)
0.1 better by far than did the exponential
extrapolation; in two sets, the data sets were not long enough beyond
R(t)=0.1 to make the distinction, and in one set,
R(t) was better fit by the exponential, but
h(t) for the same set was much better fitted by
the power law. For h(t) the results are less
secure, because the choice of the region fitted, being defined by
R(t) and not by h(t), means
that the curves are relatively noisy. For 12 curves, the power law fit
was better by far; for 4 curves, the exponential fit was as good; and
for 2 curves, the exponential fit was distinctly better. Of the 4
intermediate results, the data in 2 cases lay clearly in between the 2
extrapolations (power law and exponential) and did not fit either, and
in 2 cases, the data points were too scattered to make the
distinction.
The areas under the tail of R(t) were underestimated by the monoexponential extrapolation, with the ratio to that estimated by the power law being 0.64±0.12 (N=18). For h(t), the ratio of areas by exponential fit over that for power law was 0.63±0.08 (N=18).
From the point of view of estimation of total areas of curves, this difference is not great, a 3% to 4% underestimation for R(t) and less than that for h(t). The errors are of course greater for mean transit times, where the tails have more influence on the estimate.
Using One
for Fitting R, h, and

The most stringent test of the hypothesis is test 3. In this
approach to obtaining the best estimate of
for each experiment, the
data for R(t) and h(t) were
fit-ted simultaneously by using
(t)=a1t-
and
(t)=a2
t-
-1.
By this approach, the hypothetical fractional escape rate must
have the form
(t)=a3(a2/a1)
t-1;
the exponent is forced to be -1. The deviations of
a3 from 1.0 represent the difference in
the sensitivities of the outflow detector (for positron emissions) and
the residue detector (for 511-keV coincident gamma emissions). The best
estimates of the ai values and
were found by
nonlinear optimization using SENSOP, a sensitivity
functionbased modified Levenberg-Marquardt type routine (Chan et
al16 ). In this approach, the fractal hypothesis of a
difference of 1.0 between the power law exponents for
R(t) and h(t) is assumed,
and with this assumption having been made, the test lies in the
goodness of fit to the data for the set of three curves. Two examples
are shown in Fig 4
. The values of
so estimated are
reported in the Table
. The values of
range widely
around 2.0, averaging 2.2±0.27 (mean±SD, N=18). The constrained
theoretical lines appear to be close enough to the three data sets so
that the hypothesis cannot be rejected. The inference from both the
slopes of R(t) and of h(t)
and of the relation between their slopes is that the tails of the
transit time distributions are approximately fractal in these isolated
perfused rabbit hearts.
|
|
Generally, myocardial residue curves exhibit similarity upon scaling
time by dividing by mean transit time.2 This would be the
same as scaling by a flow-to-volume ratio or simply by
multiplying time by F when V is constant. One
might intuitively expect a higher value of
at higher flows, and
this is suggested by the positive slope of
versus F in
Fig 5
, but it is not definitive.
|
In Fig 6
, a more specific test is performed with respect
to the question of whether
is related to flow. Three sets of curves
were obtained from one heart at flows of 0.76, 1.34, and 3.44
mL · g-1 · min-1. The three residue
curves have been superimposed on each other by scaling time with
respect to mean transit time (Fig 6
, upper curves). The three outflow
curves were treated likewise (Fig 6
, lower curves). The close
juxtaposition of the curves upon each other shows that their shapes are
similar, meaning that their higher moments (variance and the skewness
and kurtosis obtained from the third and fourth moments) are close to
being the same. In Appendix A, this time scaling is applied to the
function
h(t)=A
t-
-1,
and it is shown that the exponent
is the same on the
time-scaled h(t/t) as on the
original h(t). Since scaling by flow should be
equivalent to scaling by time, the theory would predict that in one
heart at three different flows, the residue and outflow curves should
exhibit the same power law exponent
. In this experiment (the one
with the widest range of flows in the present study), the
values were 1.82, 1.91, and 2.03; ie, there is a 10% difference
between the lowest and the highest. The trend for each individual heart
in which there is a range of flows is like that for the whole data set
in Fig 5
. Thus, although the idea that
is related to flow is
neither refuted nor supported by such data, there is room for suspicion
despite the theory in Appendix A. Both Fig 5
and the analysis
of the similarity-scaled data sets of Fig 6
leave open to further
study the question of whether or not
increases with flow.
|
The absolute levels of myocardial blood flow influence the washout
curve, h(t), with higher flows giving an earlier
peak and a shorter mean transit time. The fractional escape rate of
tracer from the organ is higher at high flows. What is interesting
about the fractional escape rate is the exponent: the power law slope
of -1 is independent of the flow and of the shapes of
R(t) and h(t). This
universality is striking: whenever the system is fractal, the relation
between R(t) and h(t) gives
rise to
(t)=at-1, a power
law fractal such that in all situations the escape rate diminishes as
t-1. This is a strong statement because
(t)
is equal to the derivative of R(t) and also to
h(t) divided by 1.0 minus its integral. What this
means is that when
(t) has an exponent of -1, both
R(t) and h(t) must be power
law functions if either one is. The provocation provided by this
observation is to find the physical explanation, as discussed below,
and to ask whether it is necessarily fractal or allows some other
descriptor.
| Discussion |
|---|
|
|
|---|
It is not possible to do such experiments with standard dye dilution methods. The densitometers for indocyanine green, for example, suffer background drift of up to 2% in 1 minute with changes in blood optical density or other sources of variation, whereas with the 15O-labeled water one can measure accurately down to 10-4 times the peak values, as is seen for the outflow curves.
The interpretation of the curves is that both the residue and outflow curves demonstrate self-similarity, in the sense that for each proportional increase in time (eg, twice as long), there is a constant proportional diminution in signal (eg, one quarter as great). In the parlance of the field of nonlinear dynamics, this is termed power law behavior. Such behavior is the hallmark of fractalsthe self-similarity means that the apparent behavior (the scaling relation that says when time is twice as long, the signal is one quarter as great) is independent of the magnitude of the time unit considered.
The observation of the fractal time course of washout newly
demonstrated in the present study is made secure by the
simultaneous and coordinated measures of the three signals
R(t), h(t), and
(t). The observation that indocyanine green or
131I-albumin dilution curves were not
monoexponential but diminished more and more slowly as
time progressed has been made when recirculation was sufficiently
delayed17 or absent.18 When characterization
of washout slopes, which were concave upward on semilog plots (as in
Fig 1
, left), was important, multiexponential fits were used, for
example, for xenon washout from the brain by Hoedt-Rasmussen et
al19 ; although this was expedient and gave good estimates
of mean transit times, it also encouraged the misinterpretation that
there were regions within the organ having two or three separate flows
and inhibited the understanding that there was a broad
heterogeneity of regional flows.
Fractal Extrapolation
For the practical purpose of measuring cardiac output, Hamilton et
al13 proposed using monoexponential
extrapolation of the downslope of dilution curves to obtain an estimate
of the area under the primary first-pass indicator uncontaminated
with recirculated indicator. This was an excellent and powerful
suggestion and has been the method used ever since, even by those who
recognized that it resulted in a minor underestimation of the area
under the curve when the tails deviated (always upward) from
monoexponential. Their technique is simple and allows
an analytical calculation of the area under the extrapolated tail
beyond the section of the downslope used to estimate the exponential
rate constant. Now we propose an equally simple method of
extrapolation, but we use the power law expression to exclude the
recirculation while accounting for all the first-pass indicator and
avoiding the systematic underestimation of the exponential
extrapolation.
The method is applied to either outflow curves,
h(t), or to residue curves,
R(t). The power law exponent is the negative of
the slope of the regression of log h(t) versus
log t for the tail region where the relation is a straight
line: log h(t)=log A-ß log
t. (For the regression between times
t1 and t2, an
estimate of ß is log
[h(t1)/h(t2)]/log
[t1/t2].) The
area, area 1, up to the time t2 at which the
extrapolation begins, can be obtained directly from the data. Area 2,
under the extrapolated continuation of the observed data beyond
t2, is calculated analytically:
![]() |
![]() | (13) |
1, which
means the area extrapolated would not be finite.] The estimate of
cardiac output or flow, F, for a single path system into
which a bolus of indicator of amount qo is
injected is F=qo/(area 1+area
2).
For the residue and outflow curves in the present study, one can
calculate the degree of underestimation of the areas that would result
from using a monoexponential extrapolation. The
calculation is based on using t2 as the time
where
R(t2)/Rpeak=0.1,
and for the purpose of estimating ß, the beginning of the region
fitted was t1, the time at which
R(t1)/Rpeak=0.2.
The outflow h(t) was fitted over the identical
time period. For the 18 data sets, area 2 was calculated two ways:
first as Equation 13
, the fractal extrapolation, and second by
monoexponential extrapolation. The
monoexponential rate constant was determined from the
best fitting linear regression of log h(t) or log
R(t) versus t (linear). The results of
the two extrapolation techniques give necessarily smaller values for
the monoexponential approach: For
h(t), area 2
(monoexponential)/area 2 (fractal)=0.637±0.12 (N=18),
and for R(t), area 2
(monoexponential)/area 2 (fractal)=0.630±0.08 (N=18),
where ±1 SD is given. Naturally these errors in area due to the use of
monoexponential extrapolation are much smaller than are
errors in estimated mean transit times.
Why Is Washout Fractal?
The observed fractal washout may be explicable on the basis that
regional flows per gram of tissue have fractal spatial distributions,
as is well documented.4 5 The link between the spatial and
temporal events is not yet defined by any theory, but it makes sense
that transit times through a network with fractal flow distributions
should be fractal, since regional transit times are the local volumes
of distribution divided by the local flow.
It is interesting that the observed fractal exponents are possibly
dependent on flow. A theory for thinking that
and F
should be independent when washout curves can be superimposed on each
other by proportional time scaling2 20 is given in
Appendix A. The upward trend of
versus F in Fig 5
may be
a hint that the fractal model is imperfect, which is no doubt the case,
and some deviation toward multiexponential form is causing some degree
of flow dependency.
The observation that washout is fractal does fit with the fractal paradigm newly recognized to apply to many aspects of biology. Self-similarity over a wide range of scales is found in time-dependent functions of many sorts. As in all physical systems, the summarizing descriptor, "fractal," applies over only a finite range. Fractals are not forever, except in the mathematics of the ideal, for every real fractal relation fails at both the large and small ends of the scale. The text of Bassingthwaighte et al21 presents many examples of these self-similar scaling relations, all of them limited to finite ranges. Fractal power law scaling can emerge from any set of processes repeated multiple times with appropriate scaling between members of the set. See Appendix B for the general theory and an example using multiple exponentials.
It is intriguing that the coronary vascular network has so many fractal characteristics. The anatomy itself is fractal in terms of segment diameters and lengths,22 the number of branches is a fractal function of the diameters of the vessel from which the branches derive,23 the ratios of diameters of parent to daughter branches follow log-log scaling, and the flow distributions are spatial fractals.4 24 The degree of heterogeneity of regional flow distributions has been approximated by those calculated from artificial fractal branching patterns by using an overly simple dichotomous branching.25 Both Gan et al26 and VanBavel and Spaan8 estimate spatial distributions of flows from anatomic information; such calculations performed in our laboratory on artificial coronary systems give similar results and exhibit the same fractal correlation structure in regional flows in the heart as was observed in nature.27 The present study adds to the story but uses the completely independent evidence on washout kinetics in vivo, as opposed to structural considerations or spatial distributions of flow at a given moment of observation. Washout kinetics, quite independently of spatial patterns or branching network structures, drive us to the same conclusion that the system is fractal.
|
| Acknowledgments |
|---|
| Appendix 1 |
|---|
|
|
|---|
of the
transport functions obtained at different flows is independent of flow
can be explained mathematically on the basis of previous observations.
Similarity on scaling of the time axis by the mean transit time is
usually observed for the coronary system.2 20 28
The test of "similarity" is a statistical one; namely, can the
shape of an impulse response h1(t)
obtained at flow F1 be considered similar to the
response h2(t) obtained at flow
F2? Because there may be volume changes between
two different physiological states, rather than
scaling by
F1/F2, one
uses the more general scaling factor, the mean transit time or the
ratio of mean transit times,
1/
2,
where
i=Vi/Fi
and where the i indicates a particular state or condition or time of
day when the observation of hi(t) was
made. When similarity holds, then all time-normalized impulse
responses have the same shape and are superimposed on each other on a
plot of
ih(t/
i)
versus t/
i.
For the special situation where h(t) is a power
law function, as in Equation 11
in the text, the following equation
applies:
![]() | (14) |
i=Vi/Fi
for clarity:
![]() | (15) |
![]() |
![]() |
![]() |
)-
, and the
power law exponent
is unaffected by transformation. Since the
scaling transformation can be performed either by using
i generally or by using
Fi when Vi is constant,
then it follows that
should be unaffected by flow, as observed in
Fig 5
The same logic holds for R(t) and
(t), which have absolute values influenced by flow but
have power law exponents that are not influenced by flow.
| Appendix B |
|---|
|
|
|---|
![]() | (16) |
![]() | (17) |
The minimum mean-squared error between F(t)
and a particular f(kit)
over the interval from t=0 to t=
is found by calculating
ai:
![]() | (18) |
=kit,
substituted into Equation 18
![]() | (19) |
![]() | (20) |
A power law function can therefore be represented by a
finite sum of the scaled basis functions, where the weight of each
basis function is determined by the scale factor raised to the power
law exponent:
![]() | (21) |
In general, the ki can be chosen on the
basis of the interval over which the power law slope is fit. If the
interval is defined by t=ta to
t=tb, then
k1 can be chosen by
k1=1/ta or a conveniently
chosen value. In order to evenly distribute all of the
ki in the log-time domain, the rest of the
ki can be calculated over the range chosen:
![]() | (22) |
An example using exponentials as the basis function is
demonstrated in Fig 7
. F and f are
given as follows:
![]() | (23) |
![]() | (24) |
The finite-sum approximation is shown for N=2, 3, and 4 exponentials. An approximate fit is achieved using only four exponentials over the interval of ta=1 to tb=100. Making ta and tb outside of the desired region to be fitted and increasing N allows one to approach exact power law behavior arbitrarily closely.
Received February 21, 1995; accepted August 2, 1995.
| References |
|---|
|
|
|---|
2. Bassingthwaighte JB. Physiology and theory of tracer washout techniques for the estimation of myocardial blood flow: flow estimation from tracer washout. Prog Cardiovasc Dis. 1977;20:165-189. [Medline] [Order article via Infotrieve]
3.
Yipintsoi T, Dobbs WA Jr, Scanlon PD, Knopp TJ,
Bassingthwaighte JB. Regional distribution of diffusible
tracers and carbonized microspheres in the left ventricle of
isolated dog hearts. Circ Res. 1973;33:573-587.
4.
Bassingthwaighte JB, King RB, Roger SA.
Fractal nature of regional myocardial blood flow
heterogeneity. Circ Res. 1989;65:578-590.
5.
Glenny R, Robertson HT, Yamashiro S, Bassingthwaighte
JB. Applications of fractal analysis to
physiology. J Appl Physiol. 1991;70:2351-2367.
6.
Tomanek RJ, Searls JC, Lachenbruch PA.
Quantitative changes in the capillary bed during developing peak
and stabilized hypertrophy in the spontaneously
hypertensive rat. Circ Res. 1982;51:295-304.
7. Gonzalez F, Bassingthwaighte JB. Heterogeneities in regional volumes of distribution and flows in the rabbit heart. Am J Physiol. 1990;258(Heart Circ Physiol 27):H1012-H1024.
8.
VanBavel E, Spaan JAE. Branching patterns in
the porcine coronary arterial tree: estimation of
flow heterogeneity. Circ
Res. 1992;71:1200-1212.
9.
Richmond DR, Yipintsoi T, Coulam CM, Titus JL,
Bassingthwaighte JB. Macroaggregated albumin studies
of the coronary circulation in the dog. J
Nucl Med. 1973;14:129-134.
10. Yipintsoi T, Scanlon PD, Bassingthwaighte JB. Density and water content of dog ventricular myocardium. Proc Soc Exp Biol Med. 1972;141:1032-1035. [Medline] [Order article via Infotrieve]
11. Yipintsoi T, Tancredi R, Richmond D, Bassingthwaighte JB. Myocardial extractions of sucrose, glucose, and potassium. In: Crone C, Lassen NA, eds. Capillary Permeability (Alfred Benzon Symposium II). Copenhagen, Denmark: Munksgaard; 1970:153-156.
12.
Rose CP, Goresky CA, Bach GG. The capillary and
sarcolemmal barriers in the heart: an exploration of labeled water
permeability. Circ Res. 1977;41:515-533.
13. Hamilton WF, Moore JW, Kinsman JM, Spurling RG. Studies on the circulation, IV: further analysis of the injection method, and of changes in hemodynamics under physiological and pathological conditions. Am J Physiol. 1932;99:534-551.
14. Lerch RA, Ambos HD, Bergmann SR, Sobel BE, Ter-Pogossian MM. Kinetics of positron emitters in vivo characterized with a beta probe. Am J Physiol. 1982;242(Heart Circ Physiol 11):H62-H67.
15.
Zierler KL. Theoretical basis of
indicator-dilution methods for measuring flow and volume.
Circ Res. 1962;10:393-407.
16. Chan IS, Goldstein AA, Bassingthwaighte JB. SENSOP: A derivative-free solver for non-linear least squares with sensitivity scaling. Ann Biomed Eng. 1993;21:621-631. [Medline] [Order article via Infotrieve]
17.
Bassingthwaighte JB, Ackerman FH. Mathematical
linearity of circulatory transport. J Appl
Physiol. 1967;22:879-888.
18. Kuikka J, Levin M, Bassingthwaighte JB. Multiple tracer dilution estimates of D- and 2-deoxy-D-glucose uptake by the heart. Am J Physiol. 1986;250(Heart Circ Physiol 19):H29-H42.
19.
Hoedt-Rasmussen K, Sveinsdottir E, Lassen NA.
Regional cerebral blood flow in man determined by
intra-arterial injection of radioactive inert
gas. Circ Res. 1966;18:237-247.
20.
Yipintsoi T, Bassingthwaighte JB. Circulatory
transport of iodoantipyrine and water in the isolated dog
heart. Circ Res. 1970;27:461-477.
21. Bassingthwaighte JB, Liebovitch LS, West BJ. Fractal Physiology. New York, NY/London, England: Oxford University Press; 1994.
22. Kassab GS, Rider CA, Tang NJ, Fung YB. Morphometry of pig coronary arterial trees. Am J Physiol. 1993;265(Heart Circ Physiol 34):H350-H365.
23.
Arts T, Kruger RT, van Gerven W, Labregts JA, Reneman
RS. Propagation velocity and reflection of pressure waves in the
canine coronary artery. Am J Physiol. 1979;237:H469-H474.
24.
Bassingthwaighte JB.
Physiological heterogeneity:
fractals link determinism and randomness in structures and
functions. News Physiol Sci. 1988;3:5-10.
25. van Beek JHGM, Roger SA, Bassingthwaighte JB. Regional myocardial flow heterogeneity explained with fractal networks. Am J Physiol. 1989;257(Heart Circ Physiol 26):H1670-H1680.
26.
Gan RZ, Tian Y, Yen RT, Kassab GS. Morphometry
of the dog pulmonary venous tree. J
Appl Physiol. 1993;75:432-440.
27. Bassingthwaighte JB, Beyer RP. Fractal correlation in heterogeneous systems. Physica D. 1991;53:71-84.
28. Knopp TJ, Dobbs WA, Greenleaf JF, Bassingthwaighte JB. Transcoronary intravascular transport functions obtained via a stable deconvolution technique. Ann Biomed Eng. 1976;4:44-59.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
O. Rimoldi, K. P. Schafers, R. Boellaard, F. Turkheimer, L. Stegger, M. P. Law, A. A. Lammerstma, and P. G. Camici Quantification of Subendocardial and Subepicardial Blood Flow Using 15O-Labeled Water and PET: Experimental Validation J. Nucl. Med., January 1, 2006; 47(1): 163 - 172. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Mittal, Y. Zhou, C. Linares, S. Ung, B. Kaimovitz, S. Molloi, and G. S. Kassab Analysis of blood flow in the entire coronary arterial tree Am J Physiol Heart Circ Physiol, July 1, 2005; 289(1): H439 - H446. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. L. Munk, S. Keiding, and L. Bass Impulse-response function of splanchnic circulation with model-independent constraints: theory and experimental validation Am J Physiol Gastrointest Liver Physiol, October 1, 2003; 285(4): G671 - G680. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Staniloae, A. J. Schwab, A. Simard, R. Gallo, I. Dyrda, G. Gosselin, J. Lesperance, J. W. Ryan, and J. Dupuis In vivo measurement of coronary circulation angiotensin-converting enzyme activity in humans Am J Physiol Heart Circ Physiol, January 1, 2003; 284(1): H17 - H22. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Graham, R. K. Warrian, L. G. Girling, L. Doiron, G. R. Lefevre, M. Cheang, and W. A. C. Mutch Fractal or biologically variable delivery of cardioplegic solution prevents diastolic dysfunction after cardiopulmonary bypass J. Thorac. Cardiovasc. Surg., January 1, 2002; 123(1): 63 - 71. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. S. Kendal A stochastic model for the self-similar heterogeneity of regional organ blood flow PNAS, January 23, 2001; (2001) 21347898. [Abstract] [Full Text] |
||||
![]() |
L. M. Schwartz, T. R. Bukowski, J. D. Ploger, and J. B. Bassingthwaighte Endothelial adenosine transporter characterization in perfused guinea pig hearts Am J Physiol Heart Circ Physiol, October 1, 2000; 279(4): H1502 - H1511. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. S. Kendal A stochastic model for the self-similar heterogeneity of regional organ blood flow PNAS, January 30, 2001; 98(3): 837 - 841. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Research Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |